[1304788] The customer alleged the unit was leaking from the bottom, the liquid had a bluish tint. The unit leaked while it was powered down. The customer later stated the unit leaked cidex opa solution. The customer stated the issue was due to over-filling the reservoir with six gallons of cidex opa instead of five gallons as per the ifu (instructions for use). No injuries or adverse events were involved. The customer emptied the reservoir and refilled it with the correct amount of cidex opa solution. The issue was resolved.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2150060-2009-00158
MDR Report Key1511036
Report Source05
Date Received2009-10-13
Date of Report2009-09-18
Date of Event2009-09-18
Date Mfgr Received2009-09-18
Date Added to Maude2009-11-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactGINNY STAMBERGER
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone9497893837
Manufacturer G1MINNTECH CORP.
Manufacturer Street14605 28TH AVE. NORTH
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal Code55447
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER
Generic NameAER EQUIPMENT
Product CodeNVE
Date Received2009-10-13
Model NumberNA
Catalog Number20301
Lot NumberNA
ID NumberPART #: NA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMINNTECH CORP.
Manufacturer AddressMINNEAPOLIS MN 55447 US 55447


Patients

Patient NumberTreatmentOutcomeDate
10 2009-10-13