MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2009-10-13 for ASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER 20301 manufactured by Minntech Corp..
[1304788]
The customer alleged the unit was leaking from the bottom, the liquid had a bluish tint. The unit leaked while it was powered down. The customer later stated the unit leaked cidex opa solution. The customer stated the issue was due to over-filling the reservoir with six gallons of cidex opa instead of five gallons as per the ifu (instructions for use). No injuries or adverse events were involved. The customer emptied the reservoir and refilled it with the correct amount of cidex opa solution. The issue was resolved.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2150060-2009-00158 |
MDR Report Key | 1511036 |
Report Source | 05 |
Date Received | 2009-10-13 |
Date of Report | 2009-09-18 |
Date of Event | 2009-09-18 |
Date Mfgr Received | 2009-09-18 |
Date Added to Maude | 2009-11-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | GINNY STAMBERGER |
Manufacturer Street | 33 TECHNOLOGY DRIVE |
Manufacturer City | IRVINE CA 92618 |
Manufacturer Country | US |
Manufacturer Postal | 92618 |
Manufacturer Phone | 9497893837 |
Manufacturer G1 | MINNTECH CORP. |
Manufacturer Street | 14605 28TH AVE. NORTH |
Manufacturer City | MINNEAPOLIS MN 55447 |
Manufacturer Country | US |
Manufacturer Postal Code | 55447 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER |
Generic Name | AER EQUIPMENT |
Product Code | NVE |
Date Received | 2009-10-13 |
Model Number | NA |
Catalog Number | 20301 |
Lot Number | NA |
ID Number | PART #: NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MINNTECH CORP. |
Manufacturer Address | MINNEAPOLIS MN 55447 US 55447 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2009-10-13 |