FIBEROPTIC LIGHT CABLE 8061.456

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2009-10-16 for FIBEROPTIC LIGHT CABLE 8061.456 manufactured by Richard Wolf Medical Instruments.

Event Text Entries

[1305824] It was reported that during a laparoscopic cholecystectomy, the fiber light cable caused a 1. 000 inch x. 250 inch burn on the patient's abdomen. The appearance was of a 'bubble' without redness. Antibiotic ointment was applied to the burn.
Patient Sequence No: 1, Text Type: D, B5


[8284969] Evaluation summary - 8061. 456 - fiberoptic light cable, 4. 5mm - lot#: 184o02. The fiberoptic light cable was checked for light output and results were low light output. The adapter on the cable was loose and the ring was discolored. There were visible broken fibers with brownish color. The appearance of the cable shows wear and tear. The cable has been in use for approximately (7) seven years. The cable was returned to the customer, per their request, as is. Cause: unusual, however, we believe the broken end fibers allowed projected light to shine through the rubber insulation and heat up the rubber material.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1418479-2009-00013
MDR Report Key1517765
Report Source05,06
Date Received2009-10-16
Date of Report2009-10-16
Date of Event2009-09-17
Date Mfgr Received2009-09-23
Device Manufacturer Date2002-10-01
Date Added to Maude2010-09-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactELEANOR HALL
Manufacturer Street353 CORPORATE WOODS PARKWAY
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal60061
Manufacturer Phone8479558016
Manufacturer G1RICHARD WOLF MEDICAL INSTRUMENTS
Manufacturer Street353 CORPORATE WOODS PARKWAY
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal Code60061
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFIBEROPTIC LIGHT CABLE
Generic NameFIBEROPTIC LIGHT CABLE
Product CodeEQH
Date Received2009-10-16
Returned To Mfg2009-09-25
Model Number8061.456
Catalog Number8061.456
Lot Number184O02
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRICHARD WOLF MEDICAL INSTRUMENTS
Manufacturer AddressVERNON HILLS IL US


Patients

Patient NumberTreatmentOutcomeDate
10 2009-10-16