ASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER 20301

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2009-11-03 for ASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER 20301 manufactured by Minntech Corporation.

Event Text Entries

[16416852] The customer alleged the reservoir was leaking a mixture of cidex opa solution and water on the floor. No injuries were involved from the event. The fse assessed and repaired the unit on site. The unit is back in full operation.
Patient Sequence No: 1, Text Type: D, B5


[16423855] The vendor evaluated at the customer site. The fse stated the customer reported; "eo1" errors, system at idle. The fse inspected the system and replaced the disinfect reservoir valve due to a fluid leak. The fse also performed a test cycle while monitoring the fluid level in the disinfect reservoir. There was not an increase in the fluid level after the cycle completed. At this time, the aer unit met the manufacturer's specification.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2150060-2009-00162
MDR Report Key1521020
Report Source05
Date Received2009-11-03
Date of Report2009-10-13
Date of Event2009-10-13
Date Mfgr Received2009-10-13
Date Added to Maude2009-11-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactGINNY STAMBERGER
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone9497893837
Manufacturer G1MINNTECH CORPORATION
Manufacturer Street14605 28TH AVE. NORTH
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal Code55447
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER
Generic NameAER EQUIPMENT
Product CodeNVE
Date Received2009-11-03
Model NumberNA
Catalog Number20301
Lot NumberNA
ID NumberPART #: NA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMINNTECH CORPORATION
Manufacturer AddressMINNEAPOLIS MN 55447 US 55447


Patients

Patient NumberTreatmentOutcomeDate
10 2009-11-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.