MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2009-10-22 for ASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER 20301 manufactured by Minntech Corp..
[1303650]
The customer alleged, the unit was leaking cidex opa solution and the spray arm was not spinning. The customer stated no injuries were involved from the event. A field service engineer (fse) went to the facility, assessed the unit, and resolved the issue. The unit is currently in full operation.
Patient Sequence No: 1, Text Type: D, B5
[8361492]
The vendor evaluated at the customer's site. The fse arrived and discovered the cidex opa tank was leaking at the crack around heater element. The fse replaced the tank and verified functionality. The fse also performed test cycles without any issues. The unit met the mfr's specifications. The fse reset the timer and counter for the customer.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2150060-2009-00159 |
MDR Report Key | 1521220 |
Report Source | 05 |
Date Received | 2009-10-22 |
Date of Report | 2009-09-25 |
Date of Event | 2009-09-25 |
Date Mfgr Received | 2009-09-25 |
Date Added to Maude | 2009-11-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 0 |
Manufacturer Contact | GINNY STAMBERGER |
Manufacturer Street | 33 TECHNOLOGY DR |
Manufacturer City | IRVINE CA 92618 |
Manufacturer Country | US |
Manufacturer Postal | 92618 |
Manufacturer Phone | 9497893837 |
Manufacturer G1 | MINNTECH CORP. |
Manufacturer Street | 14605 28TH AVE. NORTH |
Manufacturer City | MINNEAPOLIS MN 55447 |
Manufacturer Country | US |
Manufacturer Postal Code | 55447 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER |
Generic Name | AER EQUIPMENT |
Product Code | NVE |
Date Received | 2009-10-22 |
Model Number | NA |
Catalog Number | 20301 |
Lot Number | NA |
ID Number | PART # NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MINNTECH CORP. |
Manufacturer Address | MINNEAPOLIS MN 55447 US 55447 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2009-10-22 |