NOVY CORNUAL CANNULATION SET J-NCS-503570

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2009-11-13 for NOVY CORNUAL CANNULATION SET J-NCS-503570 manufactured by Cook Urological Inc.

Event Text Entries

[18490431] The clear plastic tip dislodged from the blue portion during chromotubation. It was easily retrieved with a hysteroscopic grasper. These are the doctors own words. He went on to say he does approx 50-100 of these per year and would like to work on a new, but similar device with some improvements. No harm to patient.
Patient Sequence No: 1, Text Type: D, B5


[18595667] Due to the complaint device not being returned, a proper evaluation could not be performed. Two devices noting two different lot numbers wer retrieved from finished stock for evaluation. One guide catheter from the evaluation lots were test pulled and was noted to separate at the bond at 6. 9lbs, the other lot was also test pulled and was noted to separate at the bond 6. 5 lbs. Both guide catheters pulled within specification. Additional information has been requested from the physician without response at this time, when additional information is received, it will be forwarded. Unable to confirmed customers difficulty.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1825146-2009-00035
MDR Report Key1531135
Report Source99
Date Received2009-11-13
Report Date2009-11-11
Date Mfgr Received2009-10-12
Date Added to Maude2009-11-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRIS KILANDER
Manufacturer Street1100 W. MORGAN
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal47460
Manufacturer Phone8128294891
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNOVY CORNUAL CANNULATION SET
Generic NameMOV CATHETER, SALPINGOGRAPHY
Product CodeMOV
Date Received2009-11-13
Model NumberNA
Catalog NumberJ-NCS-503570
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCOOK UROLOGICAL INC
Manufacturer AddressSPENCER IN 47460 US 47460


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2009-11-13

© 2026 FDA.report