WILSON FRAME 5321G

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2009-11-12 for WILSON FRAME 5321G manufactured by Mizuho Orthopedic Systems, Inc..

Event Text Entries

[1246724] When flipping patient from supine position to prone position onto the jackson spinal table with the wilson frame added to the table. During the flip (rotation), the frame broke leaving the patient uneven on the surgical table. A new frame was installed and patient was repositioned on frame.
Patient Sequence No: 1, Text Type: D, B5


[8504716] Returned device showed evidence of staining from a chemical in the areas of the failure. We find this a unique event and will continue to monitor.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2921578-2009-00013
MDR Report Key1531405
Report Source05
Date Received2009-11-12
Date of Report2009-11-04
Date of Event2009-10-04
Date Mfgr Received2009-10-13
Device Manufacturer Date2005-04-01
Date Added to Maude2011-03-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKIRKE JAYNE
Manufacturer Street30031 AHERN AVE.
Manufacturer CityUNION CITY CA 945871234
Manufacturer CountryUS
Manufacturer Postal945871234
Manufacturer Phone5104768128
Manufacturer G1S.A.A.
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWILSON FRAME
Generic NameACCESSORIES, OPERATING ROOM TABLE
Product CodeJEB
Date Received2009-11-12
Returned To Mfg2009-09-11
Model Number5321G
Catalog Number5321G
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMIZUHO ORTHOPEDIC SYSTEMS, INC.
Manufacturer Address30031 AHERN AVE. UNION CITY CA 94587123 US 94587 1234


Patients

Patient NumberTreatmentOutcomeDate
159431. Other 2009-11-12