TI100 TRANSPORT INCUBATOR 6730070

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 1998-02-27 for TI100 TRANSPORT INCUBATOR 6730070 manufactured by Hill-rom Air-shields.

Event Text Entries

[7789123] A mfr engineer performed an eval of the device. Engineer's findings confirmed that the battery charger connector 4p1/4j1 had become overheated, which in turn caused overheating of adjacent components. A medical device safety alert was issued in 1995 notifying users of this situation. Facility indicated being aware of this notification.
Patient Sequence No: 1, Text Type: N, H10


[15261461] It was reported that a model ti100 transport incubator was in a hallway plugged into an ac receptacle to maintain constant warmth and battery charge. The unit caught fire and was subsequently extinguished by the fire department.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2510954-1998-00002
MDR Report Key153345
Report Source05,06
Date Received1998-02-27
Date of Report1998-01-29
Date Mfgr Received1998-01-29
Device Manufacturer Date1985-11-01
Date Added to Maude1998-03-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLARRY KRASLEY
Manufacturer Street330 JACKSONVILLE RD
Manufacturer CityHATBORO PA 19040
Manufacturer CountryUS
Manufacturer Postal19040
Manufacturer Phone2156828688
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTI100 TRANSPORT INCUBATOR
Generic NameTRANSPORT INCUBATOR
Product CodeFPL
Date Received1998-02-27
Returned To Mfg1998-02-01
Model NumberTI100
Catalog Number6730070
Lot Number*
ID Number*
OperatorNOT APPLICABLE
Device AvailabilityR
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key149391
ManufacturerHILL-ROM AIR-SHIELDS
Manufacturer Address330 JACKSONVILLE RD. HATBORO PA 19040 US
Baseline Brand NameTI100 TRANSPORT INCUBATOR
Baseline Generic NameTRANSPORT INCUBATOR
Baseline Model NoTI100
Baseline Catalog No6730070
Baseline IDGP00561
Baseline Device FamilyTRANSPORT INCUBATOR
Baseline Shelf Life [Months]NA
Baseline PMA FlagN
Baseline 510K PMNY
Premarket NotificationK831028
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
10 1998-02-27

© 2026 FDA.report