AFFIRM VPIII AMBIENT TEMPERATURE TRANSPORT SY 446255

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2009-11-20 for AFFIRM VPIII AMBIENT TEMPERATURE TRANSPORT SY 446255 manufactured by Becton Dickinson Infusion Therapy Systems.

Event Text Entries

[1393323] Customer was squeezing the atts ampule and a piece of glass pierced the plastic and injured the tech's finger. She was given basic first aid and sent to a workman's comp facility. They lanced her finger to remove glass that was still there and gave her antibiotics as a precaution. The tech called out of work the next day due to the pain and commented regarding a question of whether there was damage to the tendon which may require physical therapy.
Patient Sequence No: 1, Text Type: D, B5


[8336796] This product contains an ampule of reagent. The ampule is crushed and the reagent is released through a dropper into a pt sample. The customer was unable to provide a specific lot number of product which was used when this incident occurred. No further investigation can be completed with retention material. Bd offers a device called "robogrips" which customers use to crush these ampules. Further conversation with the customer revealed they are aware of the product, but never thought they were necessary because they have never had a problem before. Info regarding the robogrips has been sent to the customer for further consideration. To better educate customers regarding the use of this product, bd will be revising the package insert to add a precaution regarding this type of issue and include a reference for use of the robogrip pliers. Bd will continue to monitor this type of issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1119779-2009-00010
MDR Report Key1546587
Report Source05
Date Received2009-11-20
Date of Report2009-11-20
Date of Event2009-10-28
Date Mfgr Received2009-10-28
Date Added to Maude2011-04-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJAMES TILLMAN
Manufacturer Street7 LOVETON CIR.
Manufacturer CitySPARKS MD 21152
Manufacturer CountryUS
Manufacturer Postal21152
Manufacturer Phone4103164054
Manufacturer G1BECTON DICKINSON INFUSION THERAPY SYSTEMS
Manufacturer StreetPERIF. LUIS DONALDO COLOSIO #579, NOGALES
Manufacturer CitySONORA 84048
Manufacturer CountryMX
Manufacturer Postal Code84048
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAFFIRM VPIII AMBIENT TEMPERATURE TRANSPORT SY
Product CodeLIO
Date Received2009-11-20
Catalog Number446255
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBECTON DICKINSON INFUSION THERAPY SYSTEMS
Manufacturer AddressPERIF. LUIS DONALDO COLOSIO #579, NOGALES SONORA 84048 MX 84048


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2009-11-20

© 2026 FDA.report