MAUDE MDR 1547910

MDR report key
1547910
Report number
1418479-2009-00017
Event key
0
Event type
3
Date of event
2009-10-07
Date received
2009-11-20
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
2
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
ELEANOR HALL
Address
353 CORPORATE WOODS PKWY. VERNON HILLS IL 60061 US
Phone
847-847-8479
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1ETHMOID FORCEPS STRAIGHTETHMOID FORCEPS STRAIGHTRICHARD WOLF GMBHJXX8211.098211.09UNKY Y

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12009-11-200

Event Narratives#

D

Patient 1

DURING A SINUS PROCEDURE (LANDMARX IMAGE GUIDED FESS), THE JAW BROKE. THE JAW PIECE AND FORCEPS WERE COLLECTED FOR RETURN TO MANUFACTURER. THERE WAS NO PT CONSEQUENCE.

N

Patient 1

EVALUATION SUMMARY: 8211.09 - ETHMOID FORCEPS STRAIGHT - LOT #: UNK. THE ETHMOID FORCEPS WERE INSPECTED AND BROKEN JAW WAS CONFIRMED. THE BROKEN JAW WAS NOT INCLUDED IN THE RETURNED PACKAGE. WE DO NOT KNOW IF IT WAS DISCARDED WITHIN OUR FACILITY, OR AT THE USER FACILITY. THE ETHMOID FORCEPS IS, FROM APPEARANCE, 10 PLUS YEARS OLD. THE ETHMOID FORCEPS ARE DESIGNED TO GRASP, NOT CUT TISSUE. THE TISSUE IS REMOVED BY GRASPING AND PULLING IT GENTLY AWAY FROM THE MUSCOSA OR BONE. CAUSE: USE AND HANDLING WITH EXCESS FORCE APPLIED TO THE JAW DURING USE.