[1454454] During a sinus procedure (landmarx image guided fess), the jaw broke. The jaw piece and forceps were collected for return to manufacturer. There was no pt consequence.
Patient Sequence No: 1, Text Type: D, B5


[8501782] Evaluation summary: 8211. 09 - ethmoid forceps straight - lot #: unk. The ethmoid forceps were inspected and broken jaw was confirmed. The broken jaw was not included in the returned package. We do not know if it was discarded within our facility, or at the user facility. The ethmoid forceps is, from appearance, 10 plus years old. The ethmoid forceps are designed to grasp, not cut tissue. The tissue is removed by grasping and pulling it gently away from the muscosa or bone. Cause: use and handling with excess force applied to the jaw during use.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1418479-2009-00017
MDR Report Key1547910
Report Source06
Date Received2009-11-20
Date of Report2009-11-20
Date of Event2009-10-07
Date Facility Aware2009-10-30
Report Date2009-11-20
Date Reported to FDA2009-11-20
Date Mfgr Received2009-10-30
Date Added to Maude2010-08-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactELEANOR HALL
Manufacturer Street353 CORPORATE WOODS PKWY.
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal60061
Manufacturer Phone8479558016
Manufacturer G1RICHARD WOLF GMBH
Manufacturer StreetPFORZHEIMER STRABE 32
Manufacturer CityKNITTLENGEN 75438
Manufacturer CountryGM
Manufacturer Postal Code75438
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameETHMOID FORCEPS STRAIGHT
Generic NameETHMOID FORCEPS STRAIGHT
Product CodeJXX
Date Received2009-11-20
Returned To Mfg2009-11-06
Model Number8211.09
Catalog Number8211.09
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age10 YR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRICHARD WOLF GMBH
Manufacturer AddressKNITTLENGEN GM


Patients

Patient NumberTreatmentOutcomeDate
10 2009-11-20