ASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER 20301

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2009-11-20 for ASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER 20301 manufactured by Minntech Corporation.

Event Text Entries

[1251051] The customer alleged cidex opa solution was dripping from the scope staining the scope cabinet and the basin on the aer. The customer stated dark stains were present. The customer uses the aer unit to process the olympus scopes with model numbers h180al and qh1801al. The customer stated two scopes were processed in the aer simultaneously. The basins were not washed with any type of detergent or cleaning agent. The customer stated the cidex opa solution was clear when it dripped out of the scopes. It turned dark after making contact with the basin and scope cabinet. The customer also stated after less amount of enzol was used, the staining occurred less frequent. No patients were involved. All scopes are reprocessed if staining is witnessed.
Patient Sequence No: 1, Text Type: D, B5


[8301732] Advanced sterilization products (asp) informed the customer to use proper dilution of enzol and water for pre-cleaning and manual cleaning. Asp also informed the customer to ensure instruments are properly cleaned and rinsed per their cleaning procedures.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2150060-2009-00166
MDR Report Key1550714
Report Source05
Date Received2009-11-20
Date of Report2009-10-27
Date of Event2009-10-27
Date Mfgr Received2009-10-27
Date Added to Maude2010-01-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactGINNY STAMBERGER
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone9497893837
Manufacturer G1MINNTECH CORPORATION
Manufacturer Street14605 28TH AVE. NORTH
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal Code55447
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER
Generic NameAER EQUIPMENT
Product CodeNVE
Date Received2009-11-20
Model NumberNA
Catalog Number20301
Lot NumberNA
ID NumberPART #: NA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMINNTECH CORPORATION
Manufacturer AddressMINNEAPOLIS MN 55447 US 55447


Patients

Patient NumberTreatmentOutcomeDate
10 2009-11-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.