SIEMENS BRAND NAME: 550 TXT TABLE 7346534

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2009-11-30 for SIEMENS BRAND NAME: 550 TXT TABLE 7346534 manufactured by Siemens Medical Solutions Usa, Inc..

Event Text Entries

[1257123] A potential product issue has been reported with our oncor impression plus linear accelerator with the 550 txt table. In the abundance of caution this issue is being reported. A defective 24v power supply caused in the 550 txt table smouldering fire of the components mounted above it: relays k1, k2, k101 and the cable harness. The customer is extremely worried because of the strong smoke emission. The inner parts of the table were completely polluted by sooty particles of the burned cable insulation. Root cause is pending investigation. Corrective action is pending investigation. Risk assessement is completed. No injury or mistreatment has been reported.
Patient Sequence No: 1, Text Type: D, B5


[8303298] Severity: 3 (critical). Reason: if the potential for fire is not mitigated, it could result in serious injury or death. Physician care required to treat for smoke inhalation. Probability: preliminary c (remote) reason: smouldering fire occured for 2 out of about 430 installed tables of type 550 txt. In both cases there was no injury reported. Fire retardant components wil prevent catastrophic fire once power is removed from main source. Most of the components are also surrounded by metal. It is also reasonable to expect that the burning smell will alert the user immediately and should provide ample time to remove power from system and remove patients from the treatment area to prevent smoke inhalation or serious injury. Several interlocks are in place which would power off the system in case of short circuits caused by defective printed circuit boards. The txt treatment table mentioned and above is identified as follows: material number 7346534, (b) (4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2910081-2009-00079
MDR Report Key1559222
Report Source07
Date Received2009-11-30
Date of Report2009-11-17
Date of Event2009-11-13
Date Mfgr Received2009-11-17
Device Manufacturer Date2006-08-01
Date Added to Maude2010-02-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARK BURGESS
Manufacturer Street4040 NELSON AVE.
Manufacturer CityCONCORD CA 94520
Manufacturer CountryUS
Manufacturer Postal94520
Manufacturer Phone9256028083
Single Use3
Remedial ActionNO
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSIEMENS BRAND NAME: 550 TXT TABLE
Generic NameCOUCH, RADIATION THERAPY, POWERED
Product CodeJAI
Date Received2009-11-30
Model NumberNA
Catalog Number7346534
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS MEDICAL SOLUTIONS USA, INC.
Manufacturer Address4040 NELSON AVE. CONCORD CA 94520 US 94520


Patients

Patient NumberTreatmentOutcomeDate
10 2009-11-30