USA ELITE SYSTEM M3 AUTOCL. FOROBLIQUE SCOPE M3-30A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2009-12-17 for USA ELITE SYSTEM M3 AUTOCL. FOROBLIQUE SCOPE M3-30A manufactured by Gyrus Acmi, Inc..

Event Text Entries

[1514630] The surgeon was using an m3-gold autoclavable foroblique telescope during a turf procedure, he complained that there was no proper visualization. He was not using a camera. The surgeon aborted the case mid-way through the procedure. He went to place the catheter and realized that it was not seated properly. When he took off the drapes, he noticed that the pt's abdomen was distorted. The surgeon opened the pt and found that the bladder had been perforated. The pt spent some time in the icu, but has since been sent home. A loop and a dac cord were also used in the procedure, but were discarded by the facility.
Patient Sequence No: 1, Text Type: D, B5


[8322915] A visual examination of the scope revealed severe burn marks and excessive heat discoloration at the distal tip. The stainless steel outer tube has been severely melted in two different locations and the light fibers have been damaged at those same locations and the light fibers have been damaged at those same locations. It is also noted that moisture can be seen in the lens train using a 10x eye loupe. The distal window is chipped at 12 o'clock position when viewing from distal end and the light post up. The image is noted as cloudy due to internal damage to the telescope, likely due to overheating. Light output is at 8. 11%, well below the required 22%. Damage appears to be user error. Repair is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1519132-2009-00033
MDR Report Key1565285
Report Source07
Date Received2009-12-17
Date of Report2009-12-17
Date of Event2009-11-10
Date Facility Aware2009-11-10
Date Mfgr Received2009-11-17
Device Manufacturer Date2009-01-08
Date Added to Maude2010-09-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactMR TERRENCE SULLIVAN
Manufacturer Street136 TURNPIKE RD
Manufacturer CitySOUTHBOROUGH MA 01772
Manufacturer CountryUS
Manufacturer Postal01772
Manufacturer Phone5088042739
Manufacturer G1GYRUS ACMI, INC.
Manufacturer Street93 NORTH PLEASANT ST.
Manufacturer CityNORWALK OH 44857
Manufacturer CountryUS
Manufacturer Postal Code44857
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUSA ELITE SYSTEM M3 AUTOCL. FOROBLIQUE SCOPE
Generic NameAUTOCLAVABLE FOROBLIQUE TELESCOPE
Product CodeFBP
Date Received2009-12-17
Returned To Mfg2009-12-14
Model NumberM3-30A
Catalog NumberM3-30A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age11 MO
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGYRUS ACMI, INC.
Manufacturer Address93 NORTH PLEASANT ST. NORWALK OH 44857 US 44857


Patients

Patient NumberTreatmentOutcomeDate
10 2009-12-17