UNIVERSAL FLAT PANEL YOKE, NO CABLES, LANDSCAPE 0682400049

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00 report with the FDA on 2009-12-17 for UNIVERSAL FLAT PANEL YOKE, NO CABLES, LANDSCAPE 0682400049 manufactured by Stryker Communications.

Event Text Entries

[1273476] It was reported that within operating room #(b) (6), the monitor #1 yoke is missing the m3 screw. The issue was found prior to cases and no pts were present. No adverse consequences occurred as a result of this malfunction.
Patient Sequence No: 1, Text Type: D, B5


[8320068] There is no established expiration date for the mentioned device. Device manufactured date is unk at this point. Further attempts will be made for this info. Evaluation summary - the said device was evaluated in the field. When the stryker engineers inspected the single flat panel suspension, they noticed that the m3 screw was missing on the safety segment for the monitor. This can potentially lead to an adverse event if several cascading events occur. The safety sleeve would have to slide up the collar of the suspension and then the safety clip would have to fall out. If this took place, the monitor could fall slightly, but the cables to the monitor would likely hold the monitor up. The cause of the m3 screw missing is unk at this point. Further investigation is ongoing. This is not a single-use device.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2031963-2009-00050
MDR Report Key1566732
Report Source00
Date Received2009-12-17
Date of Report2009-11-19
Date of Event2009-11-19
Date Mfgr Received2009-11-19
Date Added to Maude2010-07-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKIM GRAUL
Manufacturer Street1410 LAKESIDE PKWY
Manufacturer CityFLOWER MOUND TX 75028
Manufacturer CountryUS
Manufacturer Postal75028
Manufacturer Phone9724107100
Manufacturer G1STRYKER COMMUNICATIONS
Manufacturer Street1410 LAKESIDE PKWY
Manufacturer CityFLOWER MOUND TX 75028
Manufacturer CountryUS
Manufacturer Postal Code75028
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameUNIVERSAL FLAT PANEL YOKE, NO CABLES, LANDSCAPE
Generic NameYOKE
Product CodeLMD
Date Received2009-12-17
Returned To Mfg2009-11-19
Model NumberNA
Catalog Number0682400049
Lot NumberNA
ID NumberNA
OperatorOTHER
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER COMMUNICATIONS
Manufacturer AddressFLOWER MOUND TX 75028 US 75028


Patients

Patient NumberTreatmentOutcomeDate
10 2009-12-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.