MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2009-12-22 for XIMATRON H72-SIM manufactured by Varian Medical Systems.
[1418861]
Customer reports that symmetric/asymmetric properties of field are changing without intervention by the radiation therapists. As this change in symmetry can alter the field size, if it is not observed by the rtt's/physicist prior to initiation of treatment, an incorrect field size could be used for treatment. Based on the report, no pt injuries or mistreatment's have occurred.
Patient Sequence No: 1, Text Type: D, B5
[8514109]
Though still under investigation, varian has determined that a mdr is appropriate, as this malfunction, should it recur, could potentially cause a serious injury. Add'l f/u to this mdr is expected upon completion of the investigation.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 8020711-2009-00009 |
MDR Report Key | 1579022 |
Report Source | 05,06 |
Date Received | 2009-12-22 |
Date of Report | 2009-11-17 |
Date of Event | 2009-11-17 |
Date Mfgr Received | 2009-11-17 |
Device Manufacturer Date | 2000-10-01 |
Date Added to Maude | 2010-02-04 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | DAVID HALL, MGR. |
Manufacturer Street | 911 HANSEN WAY |
Manufacturer City | PALO ALTO CA 94304 |
Manufacturer Country | US |
Manufacturer Postal | 94304 |
Manufacturer Phone | 6504833153 |
Manufacturer G1 | VARIAN MEDICAL SYSTEMS UK LTD. |
Manufacturer Street | GATWICK RD. |
Manufacturer City | CRAWLEY, WEST SUSSEX RH109RG |
Manufacturer Country | UK |
Manufacturer Postal Code | RH10 9RG |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | XIMATRON |
Generic Name | SYSTEM, SIMULATION, RADIATION THERAPY |
Product Code | KPQ |
Date Received | 2009-12-22 |
Model Number | H72-SIM |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | VARIAN MEDICAL SYSTEMS |
Manufacturer Address | PALO ALTO CA US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2009-12-22 |