ELECSYS 2010 RACK 11804014001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2010-02-22 for ELECSYS 2010 RACK 11804014001 manufactured by Roche Diagnostics.

Event Text Entries

[1462687] The user stated they ran one patient serum sample for free t4 and received a result of 5. 75 ng/dl which was reported. The doctor questioned the result and the same sample was repeated on (b) (6) 2010 with a result of 0. 964 ng/dl. The user did not know if the patient received any treatment based on the initial result. The free t4 reagent lot number was 15567403. The field service representative could not find a cause. He checked the sr and sipper probe adjustments, checked the liquid level detection voltages, checked and cleaned the gripper. To verify the analyzer operation, he ran performance tests, calibration, qc and precision testing with results within specifications.
Patient Sequence No: 1, Text Type: D, B5


[8440491] It was unknown if the initial reporter sent report to the fda.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2010-01146
MDR Report Key1611180
Report Source05,06
Date Received2010-02-22
Date of Report2010-02-22
Date of Event2010-01-25
Date Mfgr Received2010-01-28
Date Added to Maude2010-02-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA ERIC KOLODZIEJ
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175212834
Manufacturer G1HITACHI HIGH TECH. CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS 2010 RACK
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeKLI
Date Received2010-02-22
Model NumberNA
Catalog Number11804014001
Lot NumberNA
ID NumberNA
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0