OLYMPUS EVIS EXERA COLONOVIDEOSCOPE CF-Q160AL NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2010-02-25 for OLYMPUS EVIS EXERA COLONOVIDEOSCOPE CF-Q160AL NA manufactured by Olympus Medical Systems Corporation.

Event Text Entries

[1285636] The user facility reported to have experienced a complete loss of image during a therapeutic colonoscopy. The users reported to have utilized a different, but similar device to complete the procedure. There was no pt injury associated with this event.
Patient Sequence No: 1, Text Type: D, B5


[8449274] The device referenced in this report was returned to olympus for eval. The eval confirmed the users' report of a complete loss of image. There was evidence of fluid inside the endoscope connector and control body unit. Additionally, there was no id data transmission from the device, and the light guide tube was buckled. The cause of the users' experience was determined to be due to user mishandling. The device has been refurbished. This report is being submitted as a medical device report in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8010047-2010-00031
MDR Report Key1619424
Report Source05,06
Date Received2010-02-25
Date of Report2010-01-27
Date of Event2010-01-27
Date Mfgr Received2010-01-27
Date Added to Maude2010-08-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVE.
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS MEDICAL SYSTEMS CORPORATION
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI 192-8507
Manufacturer CountryJA
Manufacturer Postal Code192-8507
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS EVIS EXERA COLONOVIDEOSCOPE
Generic NameCOLONOSCOPE
Product CodeFTJ
Date Received2010-02-25
Returned To Mfg2010-01-28
Model NumberCF-Q160AL
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEMS CORPORATION
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI 192-8507 JA 192-8507


Patients

Patient NumberTreatmentOutcomeDate
10 2010-02-25