A200SP ABSORBER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01 report with the FDA on 2010-03-02 for A200SP ABSORBER manufactured by Penlon Ltd..

Event Text Entries

[1457630] It was reported that a (b) (6) man died after an emergency craniectomy following a road traffic accident at the (b) (6) trauma centre, (b) (6). A routine check of the a200sp absorber was carried out prior to the use. The patient's vital signs were being monitored and were within normal limits. Twenty minutes later, the fico2 increased from 0 to 15mmgh and etco2 increased from 34 to 50mmhg. The soda lime in the canister had not changed color and the unidirectional valves were functioning normally. The flows were then increased and manual ventilation was initiated. On closer inspection of canister, one of the three rubber spacers on which the inner chamber of the soda lime canister is seated was found to be missing. As a consequence the patient had been rebreathing co2. Once the spacer was replaced the fico2 and etco2 normalized. The hospital acknowledged the missing spacer was probably been due to human error when the canister was cleaned or refilled with soda lime. The instructions for use for this device state the user should check the three spacers are in place before use. Our distributor has confirmed that they provided training for the staff at the hospital. The hospital do not wish to return the product.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610531-2010-00001
MDR Report Key1619837
Report Source01
Date Received2010-03-02
Date of Report2010-01-03
Date Mfgr Received2010-01-29
Date Added to Maude2010-03-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactANGELA DUNFORD, MANAGER
Manufacturer StreetBARTON LANE ABINGDON BUSINESS PARK
Manufacturer CityABINGDON OX143PH
Manufacturer CountryUK
Manufacturer PostalOX14 3PH
Manufacturer Phone235547000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameA200SP ABSORBER
Generic NameABSORBER CARBON DIOXIDE
Product CodeBSF
Date Received2010-03-02
Model NumberNA
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPENLON LTD.
Manufacturer AddressBARTON RD ABINGDON BUSINESS PARK ABINGDON OX143PH UK OX14 3PH


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2010-03-02

© 2026 FDA.report