VALTRAC ANASTAMOSIS RING 8082-00

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 1998-04-09 for VALTRAC ANASTAMOSIS RING 8082-00 manufactured by Sherwood Davis & Geck.

Event Text Entries

[7767105] Reference mfr. Complaint # da1998-3-10-82. This complaint was deemed reportable on 3/13/1998. The actual sample was returned and evaluated. No defects were observed. The device was received in the closed position, fully engaged with all points fully interlocked. A review of the lot history was unremarkable. No previous complaints have been received against this lot since its release. Co is unable to confirm the reported problem. In fact, the product was returned in the closed position, which reportedly was the problem (device wouldn't close).
Patient Sequence No: 1, Text Type: N, H10


[15675837] Customer reports they could not get the valtrac device to close during surgery. Replacement was necessary and there was a delay of only 5-10 minutes. There was no pt complication due to this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1210157-1998-00002
MDR Report Key161994
Report Source05,06
Date Received1998-04-09
Date of Report1998-03-10
Report Date1998-03-10
Date Reported to Mfgr1998-03-10
Date Mfgr Received1998-03-10
Device Manufacturer Date1997-06-01
Date Added to Maude1998-04-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVALTRAC ANASTAMOSIS RING
Generic NameBIOFRAGMENTABLE ANASTAMOSIS RING
Product CodeLNN
Date Received1998-04-09
Returned To Mfg1998-05-05
Model NumberNA
Catalog Number8082-00
Lot Number897019
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Age9 MO
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key157730
ManufacturerSHERWOOD DAVIS & GECK
Manufacturer AddressONE CASPER ST. DANBURY CT 06810 US
Baseline Brand NameVALTRAC ANASTMOSIS RING
Baseline Generic NameANASTAMOSIS RING
Baseline Model NoNA
Baseline Catalog No8082-00
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 1998-04-09