HEARTPAK RECHARGEABLE BATTERIES 2025

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2010-02-17 for HEARTPAK RECHARGEABLE BATTERIES 2025 manufactured by Thoratec Corp..

Event Text Entries

[1334342] The pt was implanted with a left ventricular assist device (lvad). The circulatory support program coordinator reported that the pt charged the batteries overnight, went into the basement to get laundry and the batteries failed. The pt claimed that he alarmed and didn't have power to his lvad and nearly passed out. A warranty replacement set of batteries was shipped overnight to the pt at the time of the event.
Patient Sequence No: 1, Text Type: D, B5


[8490196] The pt remains ongoing on lvad support with no further issues reported. The batteries were returned to the manufacturer and are currently being analyzed. A supplemental report will be submitted when the device analysis is completed. No further info is available at this time. Additional lot# mj2250.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916596-2010-00040
MDR Report Key1626101
Report Source05,06
Date Received2010-02-17
Date of Report2010-01-19
Date of Event2010-01-18
Date Facility Aware2010-01-18
Report Date2010-01-19
Date Reported to Mfgr2010-01-19
Date Mfgr Received2010-01-19
Device Manufacturer Date2009-08-24
Date Added to Maude2010-09-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactROBERT FRYC
Manufacturer Street23 FOURTH AVE.
Manufacturer CityBURLINGTON MA 01803
Manufacturer CountryUS
Manufacturer Postal01803
Manufacturer Phone7812720139
Manufacturer G1THORATEC CORP.
Manufacturer Street6035 STONERIDGE DR.
Manufacturer CityPLEASANTON CA 94588
Manufacturer CountryUS
Manufacturer Postal Code94588
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHEARTPAK RECHARGEABLE BATTERIES
Generic NameRECHARGEABLE BATTERIES
Product CodeMOY
Date Received2010-02-17
Returned To Mfg2010-02-03
Model Number2025
Catalog Number2025
Lot NumberMJ225185
ID NumberNA
Device Expiration Date2012-07-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age5 MO
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerTHORATEC CORP.
Manufacturer AddressPLEASANTON CA US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2010-02-17