[1441496] A customer reported to beckman coulter inc. , (bci) that three patient samples gave higher than expected results using the dsl inhibin a methodology. The results are discordant to the oxford-bioinnovation inhibin assay which gave negative results. Based on available information, patients underwent further diagnostic testing.
Patient Sequence No: 1, Text Type: D, B5


[8489814] Patient samples were sent to customer product line support (cpls) for further testing. Cpls confirmed the results were due to heterophile interference. Per product labeling, the assay is not validated for use to determine gonadal maturity in children. For assays employing antibodies, the possibility exists for interference by heterophile antibodies in the patient sample.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1628193-2010-00001
MDR Report Key1632767
Report Source05
Date Received2010-03-16
Date of Report2010-03-16
Date of Event2010-02-17
Date Mfgr Received2010-02-17
Date Added to Maude2010-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street445 WEST MEDICAL CENTER BLVD.
Manufacturer CityWEBSTER TX 77598
Manufacturer CountryUS
Manufacturer Postal Code77598
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameACTIVE? INHIBIN A ELISA, 96 WELLS
Generic NameINHIBIN A ENZYME-LINKED IMMUNOSORBENT (ELISA)
Product CodeNDR
Date Received2010-03-16
Model NumberNA
Catalog NumberDSL-10-28100-1
Lot NumberUNKNOWN
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address445 WEST MEDICAL CENTER BLVD. WEBSTER TX 77598 US 77598


Patients

Patient NumberTreatmentOutcomeDate
10 2010-03-16