URINARY/CSF PROTEIN OSR6170

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-03-22 for URINARY/CSF PROTEIN OSR6170 manufactured by Beckman Coulter Biomedical Ireland Limited.

Event Text Entries

[1290308] A customer contacted beckman coulter inc. (bci) regarding negative results for urine protein patient samples generated on au chemistry system using urinary/csf protein reagent. The results were reported out of the lab. Unknown if patient treatment was affected in connection with this event.
Patient Sequence No: 1, Text Type: D, B5


[8445057] Calibration and qc were recovering with specifications. The specimen type was urine. Patient samples were requested from the customer and tested in house on the au 640 instrument and on a different analyzer. The au 640 analyzer generated negative results (except for one sample), and the different instrument gave "oir lo" results for all samples. Although reaction monitor data is consistent with an interference, a clear root cause for this event has not been determined.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2010-00125
MDR Report Key1636886
Report Source05
Date Received2010-03-22
Date of Report2010-03-22
Date of Event2010-02-22
Date Mfgr Received2010-02-23
Date Added to Maude2010-08-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER BIOMEDICAL IRELAND LIMITED
Manufacturer StreetLISMEEHAN, O'CALLAGHANS MILLS COUNTY CLARE
Manufacturer CityCOUNTY CLARE,
Manufacturer CountryEI
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameURINARY/CSF PROTEIN
Generic NameURINARY/CFS PROTEIN
Product CodeJGQ
Date Received2010-03-22
Catalog NumberOSR6170
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER BIOMEDICAL IRELAND LIMITED
Manufacturer AddressLISMEEHAN, O'CALLAGHANS MILLS COUNTY CLAIRE, EI


Patients

Patient NumberTreatmentOutcomeDate
10 2010-03-22