TRIAGE TOX DRUG SCREEN WITH MTD 94400

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2010-03-26 for TRIAGE TOX DRUG SCREEN WITH MTD 94400 manufactured by Biosite Incorporated.

Event Text Entries

[19246450] Customer reported a false negative apap result on triage tox drug screen with mtd. Customer ran four tests on one pt urine sample and obtained the following results: w45876b: +apap fresh urine. W46864b: -apap urine was cold after refrigeration. W45876b: +apap urine at room temp > 30 min. W46864b: +apap urine at room temp > 30 min. Customer used two different strip lots and had the same results on three different meters (42699, 40679 and 47056). Qc samples tested on both lots of strips with correct results. +apap is reported for this pt. Tech support checked with the tech who ran the test and she ran the urine right from the refrigerator (the negative result). She waited about a half hour and repeated, all results verified.
Patient Sequence No: 1, Text Type: D, B5


[19344494] Investigation pending.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027969-2010-00365
MDR Report Key1648917
Report Source05,06
Date Received2010-03-26
Date of Report2010-03-26
Date of Event2010-03-03
Date Mfgr Received2010-03-03
Date Added to Maude2010-05-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCARMEN BERGELIN, MANAGER
Manufacturer Street9975 SUMMERS RIDGE RD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052256
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIAGE TOX DRUG SCREEN WITH MTD
Generic NameTOX DRUG SCREEN
Product CodeDJR
Date Received2010-03-26
Model Number94400
Catalog Number94400
Lot NumberW46864B
ID NumberW45876B
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOSITE INCORPORATED
Manufacturer AddressSAN DIEGO CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2010-03-26