AMSCO AMSCO/3080

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2010-03-31 for AMSCO AMSCO/3080 manufactured by Medical Equipment Consultants, Inc..

Event Text Entries

[1297970] Patient fell from the table during surgical procedure. Table tilted deep trendelenburg to low position and head of bed 17 inches from the floor. Patient slipped from under thigh security belt onto the floor. Investigation ongoing as to if bed had any malfunction issues. Questioning new padding as more slippery less resistance with sheet and patient on bed. Noted one pad had no velcro on contact side of bed. Diagnosis or reason for use: surgery. Event abated after use stopped or dose reduced? : yes.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5015394
MDR Report Key1651064
Date Received2010-03-31
Date of Report2010-03-31
Date of Event2010-03-05
Date Added to Maude2010-04-12
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameAMSCO
Generic NameSURGICAL TABLE
Product CodeJEB
Date Received2010-03-31
Model NumberAMSCO/3080
Lot Number36051123
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerMEDICAL EQUIPMENT CONSULTANTS, INC.
Manufacturer Address12 EAGLE DR. MINSTER OH 45865 US 45865


Patients

Patient NumberTreatmentOutcomeDate
10 2010-03-31

© 2024 FDA.report