RESEARCH CRYOSTAT LEICA CM3050 S CM3050S 14047033510

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2010-04-05 for RESEARCH CRYOSTAT LEICA CM3050 S CM3050S 14047033510 manufactured by Leica Biosystems Nussloch Gmbh.

Event Text Entries

[1645369] On march 4, 2010, leica microsystems received a complaint that while using a cm3050 cryostat, a lab technician experienced a cut on her finger that required stitches. The cutting blade on this model of cryostat can be actuated by hand or by using a footswitch. The technician was not using the footswitch at the time and the footswitch was off to the side of the cryostat. Another tech approached her, looking over her shoulder to see what she was doing. The second technician inadvertently stepped on the foot switch, actuating the blade while the user's hand was in the cryostat.
Patient Sequence No: 1, Text Type: D, B5


[8595964] Conclusion from investigation: no corrective action is required. Cause of problem was due to user error - user not following instructions in user manual. Reference page 9 and 10 of current operators manual requiring that the handwheel always be "locked" in a safe position before placing hands in cryostat.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8010478-2010-00001
MDR Report Key1652714
Report Source08
Date Received2010-04-05
Date of Report2010-03-04
Date of Event2010-02-22
Date Mfgr Received2010-03-04
Device Manufacturer Date2009-09-01
Date Added to Maude2010-04-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactKLAUS STADTLER
Manufacturer StreetHEIDELBERGER STR. 17-19
Manufacturer CityNUSSLOCH 6922
Manufacturer CountryGM
Manufacturer Postal6922
Manufacturer Phone224143248
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRESEARCH CRYOSTAT LEICA CM3050 S
Generic NameCRYOSTAT
Product CodeIDP
Date Received2010-04-05
Model NumberCM3050S
Catalog Number14047033510
Lot NumberNA
ID NumberNA
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLEICA BIOSYSTEMS NUSSLOCH GMBH
Manufacturer AddressHEIDELBERGER STR. 17-19 NUSSLOCH 6922 GM 6922


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2010-04-05