ROADRUNNER TI 100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1994-09-30 for ROADRUNNER TI 100 manufactured by Air-shields Vickers, Inc..

Event Text Entries

[15484734] During transport, dc power was lost. Of the four clips on the two batteries, three clips wre found to be disconnected. Rptr found that the clips' opening had widened and loosened. Rptr checked two other incubators and found that all the clips on one were very loose and offered no resistance to removal. Rptr believes that the clips used may be inadequate.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1003557
MDR Report Key16532
Date Received1994-09-30
Date of Report1994-09-23
Date of Event1994-06-01
Date Added to Maude1994-10-03
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameROADRUNNER
Generic NameTRANSPORT INCUBATOR
Product CodeFPL
Date Received1994-09-30
Model NumberTI 100
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key16465
ManufacturerAIR-SHIELDS VICKERS, INC.
Manufacturer AddressHATBORO PA 19040 US


Patients

Patient NumberTreatmentOutcomeDate
10 1994-09-30