MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00 report with the FDA on 2010-04-12 for CLIN CHEM ALBUMIN BCG REAGENT 7D53-21 manufactured by Abbott Manufacturing, Inc..
[1579557]
The customer observed inconsistent clinical chemistry albumin bcg quality control (qc) values when reagent lot 80051hw00 and 83030hw00 were in use. The customer stated the first few albumin calibrations failed, therefore, the technician re-mixed the reagent and calibrators and calibration was successful. The customer performed the recommended troubleshooting procedures and verified the correct albumin assay version was in use. The customer monitored the issue and called back to state that since the techs were mixing the reagent, the albumin calibrations were passing and the qc was within range. There was no impact to patient management.
Patient Sequence No: 1, Text Type: D, B5
[8541673]
(b)(4). Suspect medical device, lot #: additional clinical chemistry albumin bcg reagent lot: 83030hw00, expiration date: 10/10/11. Concomitant medical products: architect c8000 analyzer, list # 1g06-11, (b)(4). This is an initial report. An investigation is in process. A final report will be submitted when the investigation is complete.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2018433-2010-00063 |
MDR Report Key | 1656649 |
Report Source | 00 |
Date Received | 2010-04-12 |
Date of Report | 2010-02-16 |
Date of Event | 2010-02-16 |
Date Mfgr Received | 2010-04-14 |
Date Added to Maude | 2010-12-10 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | NOEMI KONDOS RN, BSN |
Manufacturer Street | 200 ABBOTT PARK ROAD DEPT. 09B9, AP50 |
Manufacturer City | ABBOTT PARK IL 600643537 |
Manufacturer Country | US |
Manufacturer Postal | 600643537 |
Manufacturer Phone | 8479375120 |
Manufacturer G1 | ABBOTT MANUFACTURING, INC. |
Manufacturer Street | 820 MISSION ST. |
Manufacturer City | SOUTH PASADENA CA 91030 |
Manufacturer Country | US |
Manufacturer Postal Code | 91030 |
Single Use | 3 |
Remedial Action | RC |
Previous Use Code | 3 |
Removal Correction Number | 2018433-3/11/10-001-R |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CLIN CHEM ALBUMIN BCG REAGENT |
Generic Name | QUANTITATION OF ALBUMIN IN HUMAN SERUM OR PLASMA |
Product Code | CIX |
Date Received | 2010-04-12 |
Catalog Number | 7D53-21 |
Lot Number | 80051HW00 |
Device Expiration Date | 2010-07-15 |
Operator | OTHER |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT MANUFACTURING, INC. |
Manufacturer Address | 820 MISSION ST. SOUTH PASADENA CA 91030 US 91030 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2010-04-12 |