SKINTACT WR01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,08 report with the FDA on 2010-04-29 for SKINTACT WR01 manufactured by Leonhard Lang Gmbh.

Event Text Entries

[1525197] On 03/25, we have been informed about an injury of a pt at a hospital in (b) (6). The date and nature of the surgery are still unk to us. A non-monitoring dispersive electrode (model wr01) and a electrosurgical generator (no model has been revealed) were used. No info of the size and the degree of the burn, how it was treated afterwards, how the skin was prepared, the orientation of the electrode, the generator model, the power setting used could be obtained so far.
Patient Sequence No: 1, Text Type: D, B5


[8530244] The device involved in the incident and 4 other electrodes were returned in their original pouch. The device involved, two of the returned samples of the same lot and retained samples of the same lot have been tested visually, electrically, and thermally. Mechanical test was performed on 2 returned and on 2 retained samples (the device involved in the incident was in a state not suitable for the mechanical test). All devices were found to perform within limits. No faults could be detected. The involved electrode shows no stains or marks that would indicate a burn. We are therefore, not sure, whether this incident is reportable at all. No conclusion can be drawn so far. We will continue to ask for further info and will relay such info and any conclusion in a f/u report.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8020045-2010-00007
MDR Report Key1672801
Report Source01,08
Date Received2010-04-29
Date of Report2010-04-22
Date Mfgr Received2010-03-25
Device Manufacturer Date2008-04-01
Date Added to Maude2010-05-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. BURRHUS LANG
Manufacturer StreetARCHENWEG 56
Manufacturer CityINNSBRUCK 6020
Manufacturer CountryAU
Manufacturer Postal6020
Manufacturer Phone12334257
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSKINTACT
Generic NameELECTROSURGICAL GROUNDING PLATE
Product CodeODR
Date Received2010-04-29
Returned To Mfg2010-04-01
Model NumberWR01
Lot Number80408-0803
Device Expiration Date2010-04-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLEONHARD LANG GMBH
Manufacturer AddressARCHENWEG 56 INNSBRUCK 6020 AU 6020


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2010-04-29

© 2026 FDA.report