CLIN CHEM ALBUMIN BCG REAGENT 7D53-21

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00 report with the FDA on 2010-05-04 for CLIN CHEM ALBUMIN BCG REAGENT 7D53-21 manufactured by Abbott Manufacturing, Inc..

Event Text Entries

[1489365] The customer observed clinical chemistry albumin bcg quality control results below peer range after switching to the reformulated albumin reagents. The customer did not observe this issue with other assays. The customer provided precision data for evaluation and also provided the following example of patient results as follows: original reagent: 4. 14 g/dl, reformulated reagent 4. 37 g/dl. The customer also performed some recommended troubleshooting and was asked to monitor the issue. There was no impact to patient management.
Patient Sequence No: 1, Text Type: D, B5


[8570999] (b)(4), concomitant medical products:architect c8000 analyzer, list # 1g06-01, serial # (b)(4),this is an initial report. An investigation is in process. A final report will be submitted when the investigation is complete.
Patient Sequence No: 1, Text Type: N, H10


[8860238] (b)(4). An incorrect suspect medical device, lot number, expiration date was submitted in the initial medwatch submission. The expiration date has been corrected to 10/15/11. Upon further review, the customer's issue is not associated with remedial action (b)(4), and is unassigned.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2018433-2010-00071
MDR Report Key1672832
Report Source00
Date Received2010-05-04
Date of Report2010-03-12
Date of Event2010-03-12
Date Mfgr Received2011-01-19
Date Added to Maude2011-12-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNOEMI KONDOS RN, BSN
Manufacturer Street200 ABBOTT PARK ROAD DEPT. 09B9, AP50
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone8479375120
Manufacturer G1ABBOTT MANUFACTURING, INC.
Manufacturer Street820 MISSION ST.
Manufacturer CitySOUTH PASADENA CA 91030
Manufacturer CountryUS
Manufacturer Postal Code91030
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number2018433-3/11/10-001-R
Event Type3
Type of Report3

Device Details

Brand NameCLIN CHEM ALBUMIN BCG REAGENT
Generic NameQUANTITATION OF ALBUMIN IN HUMAN SERUM OR PLASMA
Product CodeCIX
Date Received2010-05-04
Catalog Number7D53-21
Lot Number83030HW00
Device Expiration Date2011-10-15
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING, INC.
Manufacturer Address820 MISSION ST. SOUTH PASADENA CA 91030 US 91030


Patients

Patient NumberTreatmentOutcomeDate
10 2010-05-04

© 2026 FDA.report