ELECTROSURGICAL GENERATOR AND PENCIL SSE4, E2515H

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1994-10-12 for ELECTROSURGICAL GENERATOR AND PENCIL SSE4, E2515H manufactured by Valleylab, Inc..

Event Text Entries

[15254191] Pt was in surgery for bilateral blepharoplasty, under local anesthesia with nasal o2. Physician had completed the right eye and was completing the left eye using the electrosurgical machine with a pencil. Physician was using pencil on orbital fat when a loud noise was heard and a flame was noted to pt's lower face (nares chin area). Physician extinguished the fire with his hands and water was poured on pt's face. Pt was examined by plastic surgeon and burned area evaluated as 1st and 2nd degree burns. Physician ordered ice packs and cortisporin ointment to area from nares to chin.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1003661
MDR Report Key16778
Date Received1994-10-12
Date of Report1994-09-27
Date of Event1994-09-23
Date Added to Maude1994-10-14
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameELECTROSURGICAL GENERATOR AND PENCIL
Generic NameGENERATOR/HAND SWITCHING PENCIL
Product CodeHAM
Date Received1994-10-12
Model NumberSSE4, E2515H
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key16711
ManufacturerVALLEYLAB, INC.
Manufacturer AddressBOULDER CO 80301 US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 1994-10-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.