UNICEL DXC 800 SYNCHRON CHEMISTRY ANALYZER N/A A11812

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-05-27 for UNICEL DXC 800 SYNCHRON CHEMISTRY ANALYZER N/A A11812 manufactured by Beckman Coulter Inc.

Event Text Entries

[1508304] Customer contacted beckman coulter inc. (bci) in regards to several erroneous low phosphorous (phos) patient results, which were generated by unicel dxc 800 pro chemistry analyzer. One patient was treated based on the erroneous results, just before the amended result was called in. The results were reported out of the laboratory. An amended reports were issued.
Patient Sequence No: 1, Text Type: D, B5


[8626871] The samples were serum. Calibration appears acceptable before the event. Qc chart from the previous week resulted in one low and two high phosm fliers. A field service engineer (fse) replaced the phosm module and associated tubing. A clear root cause can not be identified for this event.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2010-00298
MDR Report Key1701448
Report Source05
Date Received2010-05-27
Date of Report2010-05-27
Date of Event2010-05-04
Date Mfgr Received2010-05-04
Device Manufacturer Date2009-01-22
Date Added to Maude2010-05-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613436
Manufacturer G1COULTER BECKMAN INC.
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use0
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUNICEL DXC 800 SYNCHRON CHEMISTRY ANALYZER
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeCEO
Date Received2010-05-27
Model NumberN/A
Catalog NumberA11812
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821 US 92821