GENDEX GXP DENTAL X-RAY PROCESSOR 112-0527G1 A0400

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2010-05-27 for GENDEX GXP DENTAL X-RAY PROCESSOR 112-0527G1 A0400 manufactured by Gendex Dental Systems.

Event Text Entries

[1398721] The doctor stated that during operation of the dental x-ray film processor, the dryer fan stopped operating while the heater continued to operate. This resulted in the heater assembly overheating and starting on fire. The fire was localized to the film processor.
Patient Sequence No: 1, Text Type: D, B5


[8658400] There was no property damage or user injury with this event. The top cover assembly of the film processor contains a fan and heater that operates during the drying of the film as part of the x-ray film processing. The fan creates the air flow over the heater for drying the x-ray film. The fan stopped working for unknown reasons possibly contributing to the overheating of the heater unit. The manufacturer has requested return of the fan and heater from the dental office for evaluation. The film processing unit has been repaired with a fan and heater replacement and is operational.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004115000-2010-00009
MDR Report Key1701469
Report Source06
Date Received2010-05-27
Date of Report2010-04-27
Date of Event2010-04-27
Date Mfgr Received2010-04-27
Device Manufacturer Date1996-05-27
Date Added to Maude2010-07-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DAWN BRAGG
Manufacturer Street901 WEST OAKTON STREET
Manufacturer CityDES PLAINES IL 600181884
Manufacturer CountryUS
Manufacturer Postal600181884
Manufacturer Phone8473643290
Manufacturer G1GENDEX DENTAL SYSTEMS
Manufacturer Street901 WEST OAKTON STREET
Manufacturer CityDES PLAINES IL 60018188
Manufacturer CountryUS
Manufacturer Postal Code60018 1884
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGENDEX GXP DENTAL X-RAY PROCESSOR
Generic NameDENTAL X-RAY FILM PROCESSOR
Product CodeEGY
Date Received2010-05-27
Model Number112-0527G1
Catalog NumberA0400
OperatorDENTIST
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerGENDEX DENTAL SYSTEMS
Manufacturer Address901 WEST OAKTON STREET DES PLAINES IL 60018188 US 60018 1884


Patients

Patient NumberTreatmentOutcomeDate
10 2010-05-27

© 2026 FDA.report