QUINTON TM55 TREADMILL 000489-009

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2010-05-28 for QUINTON TM55 TREADMILL 000489-009 manufactured by Cardiac Science Corporation.

Event Text Entries

[1538458] Customer stated that during a stress test, the pt requested the treadmill to be slowed down. The end user slowed the treadmill and attempted to stop the treadmill belt but it was non-responsive. The pt fell from the treadmill and was taken to the emergency room to be evaluated. The extent of the injuries is unk.
Patient Sequence No: 1, Text Type: D, B5


[8550327] Csc field service engineer was dispatched on-site to evaluate the treadmill and stress testing sys. The sys was tested several times and the problem could not be duplicated. As a precaution, the treadmill interface cable was replaced. Log files from the sys were also sent in for eval but they were incomplete and nothing that would contribute to the investigation was logged. No evidence there was a device problem, this could have been user error.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3014398-2010-00016
MDR Report Key1708444
Report Source06
Date Received2010-05-28
Date of Report2010-05-26
Date of Event2010-04-29
Date Mfgr Received2010-04-29
Device Manufacturer Date2002-12-01
Date Added to Maude2010-06-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactBEV MAGRANE
Manufacturer Street3303 MONTE VILLA PKWY.
Manufacturer CityBOTHELL WA 98028
Manufacturer CountryUS
Manufacturer Postal98028
Manufacturer Phone4254022365
Manufacturer G1CARDIAC SCIENCE CORPORATION
Manufacturer Street500 BURDICK PARKWAY
Manufacturer CityDEERFIEDL WI 53531
Manufacturer CountryUS
Manufacturer Postal Code53531
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameQUINTON TM55 TREADMILL
Generic NameCARDIAC STRESS TESTING TREADMILL
Product CodeIOL
Date Received2010-05-28
Model Number000489-009
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCARDIAC SCIENCE CORPORATION
Manufacturer Address3303 MONTE VILLA PKWY. BOTHELL WA 98028 US 98028


Patients

Patient NumberTreatmentOutcomeDate
10 2010-05-28

© 2026 FDA.report