SELECT FLOWMETER OHMEDA 3MFA1005

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2010-05-21 for SELECT FLOWMETER OHMEDA 3MFA1005 manufactured by Precision Medical.

Event Text Entries

[1396231] Flowmeter had a crack in its base which was not easily visible, but a modest weight (e. G. , humidifier) attached was enough to spread the crack and cause leakage. The flowmeter indicated proper flow, but very little of it entered the humidifier to the patient. When the problem was discovered and the flowmeter was replaced, the result was "noticeable improvement" in the patient's o2 saturation. ======================health professional's impression======================physical damage was already visible on the flowmeter when it was put in use. Most of one hose barb and a 2-cm portion of the base were broken off. Attempted use of the obviously damaged flowmeter led to the problem of the leaking crack and inadequate oxygenation of the patient, which may have been avoided if the flowmeter had been reported and sent for testing after it was damaged. ======================manufacturer response for oxygen flowmeter, select======================mfr was given details and photographs, and did not think it was necessary to have flowmeter sent to them.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1710460
MDR Report Key1710460
Date Received2010-05-21
Date of Report2010-05-13
Date of Event2010-05-12
Report Date2010-05-13
Date Reported to FDA2010-05-21
Date Added to Maude2010-06-04
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameSELECT FLOWMETER
Generic NameFLOWMETER, TUBE, THORPE, BACK-PRESSURE COMPENSATED
Product CodeCAX
Date Received2010-05-21
Model NumberOHMEDA
Catalog Number3MFA1005
Lot Number*
ID Number*
Device AvailabilityN
Device Sequence No1
Device Event Key0
ManufacturerPRECISION MEDICAL
Manufacturer Address300 HELD DR. NORTHAMPTON PA 18067 US 18067


Patients

Patient NumberTreatmentOutcomeDate
10 2010-05-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.