[1536432] Pt's mom called in and said that her son, pt: (b) (6), fell out of his standing wheelchair (model: hps-2, (b) (4)), and hit his head.
Patient Sequence No: 1, Text Type: D, B5


[8548368] Device was returned to our facility on 06/12/2008. Device was inspected. Recreation of event on (b) (6) 2008 at device mfg facility. Upon investigation, it was noted that the set screw on the arm swivel housing, which is loctited, the loctite does not adhere properly once arm assembly is installed with grease. Device was replaced with new arm assembly. Corrective action was generated from investigation. Device was returned to pt (b) (6) in (b) (6) 2008, where he was able to continue use of the chair. Issue was resolved.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3007494904-2008-00001
MDR Report Key1712829
Report Source04
Date Received2010-06-05
Date of Report2010-06-02
Date of Event2008-06-04
Date Mfgr Received2008-06-04
Device Manufacturer Date2006-10-01
Date Added to Maude2010-06-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street5848 DIXIE HIGHWAY
Manufacturer CitySAGINAW MI 486015967
Manufacturer CountryUS
Manufacturer Postal486015967
Manufacturer Phone9897469100
Manufacturer G1THE STANDING COMPANY
Manufacturer Street5848 DIXIE HIGHWAY
Manufacturer CitySAGINAW MI 48601596
Manufacturer CountryUS
Manufacturer Postal Code48601 5967
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHALF POWER SUPERSTAND STANDUP WHEELCHAIR
Generic NameSTANDUP WHEELCHAIR
Product CodeIPL
Date Received2010-06-05
Returned To Mfg2008-06-12
Model NumberHPS-2
Catalog NumberHPS-2
Lot NumberNA
ID NumberNA
OperatorLAY USER/PATIENT
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTHE STANDING COMPANY
Manufacturer Address5848 DIXIE HIGHWAY SAGINAW MI 48601 US 48601


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2010-06-05