UNIVERSAL COUCHTOP RECTANGULAR EXTENSION MTIL6600

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,08 report with the FDA on 2010-06-11 for UNIVERSAL COUCHTOP RECTANGULAR EXTENSION MTIL6600 manufactured by Medtec, Inc..

Event Text Entries

[17534288] Prone patient climbed onto the couch with his head and shoulders over the extension. He was then asked to move and grabbed the sides of the extension with his hands lifting himself up slightly. The extension then came free of the main couch resulting in the patient falling to the floor and hitting himself upon the theraview imaging unit. The patient was described as large but not the largest. The patient appeared unharmed and continued treatment on a difference machine.
Patient Sequence No: 1, Text Type: D, B5


[17642045] Results: locking mechanism. Method: distributor's description of events were reviewed. These observations described were able to be duplicated with product evaluated from inventory by mechanically distributing a load unevenly across the device. Results: clinicians did not follow instructions for using the device and allowed patients to reposition themselves without aid. Uneven loading and shifting on the device result in a force placed on the axel of the locking mechanism resulting in its rotating until it reaches the release point and the extension disengages. Conclusion: user error allowing patient to reposition unaided contributed to device failure. Remedial action initiated: notifications sent to customers identifying the potential failure method for the extension and providing reminders on how to utilize the product. New re-designed latches with positive locking feature are being sent out to all customers to replace the current latches on their extension units.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1932738-2010-00002
MDR Report Key1725408
Report Source01,08
Date Received2010-06-11
Date of Report2010-06-11
Date of Event2010-05-21
Date Mfgr Received2010-05-21
Device Manufacturer Date2009-12-01
Date Added to Maude2012-03-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street1401 8TH ST. S.E.
Manufacturer CityORANGE CITY IA 51041
Manufacturer CountryUS
Manufacturer Postal51041
Manufacturer Phone7127378688
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUNIVERSAL COUCHTOP RECTANGULAR EXTENSION
Generic NameUNIVERSAL COUCHTOP RECTANGULAR EXTENSION
Product CodeIXQ
Date Received2010-06-11
Model NumberMTIL6600
Catalog NumberMTIL6600
Lot NumberM161450
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMEDTEC, INC.
Manufacturer Address1401 8TH ST. S.E. ORANGE CITY IA 51041 US 51041


Patients

Patient NumberTreatmentOutcomeDate
10 2010-06-11

© 2024 FDA.report