STRAIGHT COLD KNIFE * 27069K

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2010-06-04 for STRAIGHT COLD KNIFE * 27069K manufactured by Karl Storz.

Event Text Entries

[1411098] The tip of the cold knife broke off during the procedure. The first time the tip was retrieved with some difficulty. This was done using a grasping forceps from the urethra via cystoscope. However, since the intended procedure was not yet complete, a second cold knife was used. The second time the tip broke off and could not be retrieved via cystoscopy. The c-arm was then used to visualize and the tip was seen in the retroperitoneal space. The surgeon had to do an exploratory laparotomy and a perineal exploration to retrieve the tip. It was found close to a major vessel but had not penetrated any vital structures. A third device was opened but not used on the patient for comparison with the fluoro images and this device broke in the same place as well.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1731769
MDR Report Key1731769
Date Received2010-06-04
Date of Report2010-06-02
Date of Event2010-06-02
Report Date2010-06-02
Date Reported to FDA2010-06-04
Date Added to Maude2010-06-22
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameSTRAIGHT COLD KNIFE
Generic NameURETHROTOME, COLD KNIFE, DISPOSABLE
Product CodeEZO
Date Received2010-06-04
Model Number*
Catalog Number27069K
Lot Number38237 ALL 3 SAME LOT #
ID Number*
OperatorPHYSICIAN
Device AvailabilityR
Device Age1 DY
Device Sequence No1
Device Event Key0
ManufacturerKARL STORZ
Manufacturer Address2151 E. GRAND AVENUE EL SEGUNDO CA 90245 US 90245


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2010-06-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.