SPACELABS ULTRAVIEW SL MULTIGAS ANALYZER MODULE 91518

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-06-03 for SPACELABS ULTRAVIEW SL MULTIGAS ANALYZER MODULE 91518 manufactured by Spacelabs Healthcare.

Event Text Entries

[1339808] Spacelabs rec'd a report that a multigas analyzer module displayed question marks rather than agent values when a single agent was delivered. The display showed "des" and "sevo" and no agent values when delivering more than one agent. The report state that the device was working when case started and then exhibited the problem well into the case. The operation of the device did not affect pt outcome.
Patient Sequence No: 1, Text Type: D, B5


[8617174] The device was returned to spacelabs (b)(4) for testing. Performance testing confirmed that the gas bench had malfunctioned. The gas bench was replaced and device operation confirmed. The customer's unit has been replaced. We have concluded that no further action is required and consider this issue closed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3023361-2010-00014
MDR Report Key1734897
Report Source05
Date Received2010-06-03
Date of Report2010-06-03
Date of Event2010-05-06
Date Mfgr Received2010-05-06
Device Manufacturer Date2009-06-01
Date Added to Maude2011-04-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDAVID GERAGHTY, MGR
Manufacturer Street5150 220TH AVE SE
Manufacturer CityISSAQUAH WA 98029
Manufacturer CountryUS
Manufacturer Postal98029
Manufacturer Phone4256577200
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSPACELABS ULTRAVIEW SL MULTIGAS ANALYZER MODULE
Generic NameANESTHESIA GAS ANALYZER MODULE
Product CodeCBR
Date Received2010-06-03
Returned To Mfg2010-05-13
Model Number91518
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSPACELABS HEALTHCARE
Manufacturer Address5150 220TH AVE SE ISSAQUAH WA 98027 US 98027


Patients

Patient NumberTreatmentOutcomeDate
10 2010-06-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.