SAF-D-FIB 1220-002 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1994-10-05 for SAF-D-FIB 1220-002 NA manufactured by Medtronic.

Event Text Entries

[22185747] During a cardiac arrest, while patient in pulseless ventricular tachycardia, an attempt was made to defibrillate the patient. The defibrillator pads were applied to the patient. A lifepack was used to defibrillate patient. The defibrillator discharged but there was no evidence that the shock went through to the patient. A second lifepack was then used with the same defibrillator pads. The defibrillator discharged but again there was no evidence that the shock went through to the patient. The defibrillator pads were then removed and pads manufactured by a different company were applied. The patient was then able to be defibrillated. There were no markings or burns noted at the defibrillator pad sites.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number17468
MDR Report Key17468
Date Received1994-10-05
Date of Report1994-06-09
Date of Event1994-05-29
Date Facility Aware1994-06-01
Report Date1994-06-09
Date Reported to Mfgr1994-06-09
Date Added to Maude1994-11-08
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameSAF-D-FIB
Generic NameDEFIBRILLATOR PAD
Product CodeDRK
Date Received1994-10-05
Model Number1220-002
Catalog NumberNA
Lot NumberUNKNOWN
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Age*
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key17396
ManufacturerMEDTRONIC
Manufacturer Address60 NEWARK STREET HAVERHILL MA 018321398 US


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening 1994-10-05