9008 ANALETTE II CHEMISTRY ANALYZER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2010-07-12 for 9008 ANALETTE II CHEMISTRY ANALYZER manufactured by Precision Systems, Inc..

Event Text Entries

[17729591] Precision systems, inc is advertising a diagnostic chemistry analyzer, the analette ii. This chemistry analyzer is an "open" system, meaning that there should be various reagent manufacturer's to choose from. The manufacturer, precision systems, inc. Is required to perform validation studies on this chemistry analyzer from either their own brand name reagents, or other reagents before the system is market ready. When i called this company to determine what reagent manufacturer's have, they validated on their instrument, they gave me "microgenics"; which is manufactured by thermoscientific. I called thermoscientific and they absolutely do not offer reagents for the analette ii chemistry analyzer. When i called precision systems, inc back and asked them what reagents did they use to validate their instrument, i spoke with (b)(4), he again stated microgenics, and another company that he could not remember the name of. I told him that i spoke with thermoscientific and their reagents were not validated on this instrument. He told me that thermoscientific had their equipment for some years and was doing validation studies, but abruptly stopped. I asked him if he had data verifying the validation for any reagents. He told me he did, and when i asked to see the data, he told me i was not privy to that. I told him i was privy to that as a consumer, and that there were several problems here and that i was going to turn his company into the fda adverse medical device program. First of all, they are selling a diagnostic medical device that has not been validated by any reagent manufacturer, and i as a consumer would not be able to purchase reagents in the united states to perform patient testing on. Second, they were not allowed by clia guidelines, and who knows what other guidelines to sell this instrument unless it is validated, and the statistics were made available to me, the consumer. I cannot validate that it has been fda approved as well. As a medical technologist for over 20 years, and a laboratory consultant for as many years, i cannot tell you how easy it is for my clients to scour the internet and find instruments of this caliber. Most of my clients do not know the important questions to ask, and who knows how many have purchased it to date. It is very scary to see.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5016720
MDR Report Key1759598
Date Received2010-07-12
Date of Report2010-07-12
Date of Event2010-07-12
Date Added to Maude2010-07-21
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand Name9008 ANALETTE II CHEMISTRY ANALYZER
Generic NameUNK
Product CodeJJF
Date Received2010-07-12
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerPRECISION SYSTEMS, INC.


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening; 2. Other 2010-07-12