HOYER LIFT W/SCALE HPL700 W/SCALE HPL700/WSC

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00 report with the FDA on 2010-07-23 for HOYER LIFT W/SCALE HPL700 W/SCALE HPL700/WSC manufactured by Sr Instruments, Inc. (scale Manufacturer Only).

Event Text Entries

[20916611] It was reported to manufacturing by the distributor via the facility. Per the facility called stating that during a pt transfer the lift broke, the bolt/pin coming up through the cradle to the scale broke, causing the assembly to tilt and then a dog ear piece broke off resulting in this fall. The resident fell back onto the bed with the cradle landing on top of her, no apparent injury, facility still accessing the incident. An ra was issued by the distributor to get the scale and cradle returned for evaluation. Said evaluation was received from the distributor.
Patient Sequence No: 1, Text Type: D, B5


[21228558] Investigation: telephone conferences were initiated between (b)(6) and sr instruments. Testing initiated trying to reproduce the mode of failure, to determine if the device was used properly or as intended. Sr instruments doing material testing, verified steels heat treating processes and verified material type. (b)(4) performing safety performance testing.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1317256-2010-00001
MDR Report Key1780900
Report Source00
Date Received2010-07-23
Date of Report2010-07-22
Date of Event2010-04-13
Date Mfgr Received2010-05-11
Device Manufacturer Date2010-03-01
Date Added to Maude2012-03-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street600 YOUNG ST.
Manufacturer CityTONAWANDA NY 14150
Manufacturer CountryUS
Manufacturer Postal14150
Manufacturer Phone7166935977
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHOYER LIFT W/SCALE
Generic NamePATIENT LIFTER W/SCALE
Product CodeFRW
Date Received2010-07-23
Returned To Mfg2010-05-19
Model NumberHPL700 W/SCALE
Catalog NumberHPL700/WSC
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSR INSTRUMENTS, INC. (SCALE MANUFACTURER ONLY)
Manufacturer Address600 YOUNG ST TONAWANDA NY 14150 US 14150


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2010-07-23

© 2026 FDA.report