UNICEL? DXL 800 ACCESS? IMMUNOASSAY SYSTEM DXI 800 973100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-08-03 for UNICEL? DXL 800 ACCESS? IMMUNOASSAY SYSTEM DXI 800 973100 manufactured by Beckman Coulter Inc..

Event Text Entries

[1628202] A customer contacted beckman coulter inc. (bci) regarding a ckmb result above the normal reference range generated by the unicel dxi 800 access immunoassay system for one patient. Upon repeat, the result was within the normal reference range. No reports of death, injury, or change to patient treatment have been reported in connection with this event.
Patient Sequence No: 1, Text Type: D, B5


[8728385] Sample was li heparin plasma collected in a pst tube and centrifuged for 5 minutes at 6000 rpm. Qc was within the customer's established ranges prior to and after the event. A system check performed on 06/14/2010 met specifications. Per customer, this sample collection tube was only 2/3 of the recommended fill volume. A field service engineer (fse) went on-site and replaced some hardware after which a system check, precision run and qc performed all met specifications. Although a sample issue may have contributed to this event, a clear root cause for this event has not been determined to date.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2122870-2010-00388
MDR Report Key1781561
Report Source05
Date Received2010-08-03
Date of Report2010-07-30
Date of Event2010-07-02
Date Mfgr Received2010-07-02
Device Manufacturer Date2007-08-10
Date Added to Maude2010-08-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BOULEVARD
Manufacturer CityBREA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street1000 LAKE HAZELTINE DRIVE
Manufacturer CityCHASKA MN 55318
Manufacturer CountryUS
Manufacturer Postal Code55318
Single Use3
Remedial ActionIN
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameUNICEL? DXL 800 ACCESS? IMMUNOASSAY SYSTEM
Generic NameDISCRETE PHOTOMETRIC CHEMISTRY ANALYZER
Product CodeJHS
Date Received2010-08-03
Model NumberDXI 800
Catalog Number973100
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address1000 LAKE HAZELTINE DRIVE CHASKA MN 55318 US 55318


Patients

Patient NumberTreatmentOutcomeDate
10 2010-08-03

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.