ZOOM WHITENING LAMP AND PROCEDURE KIT ZM2550

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04,05 report with the FDA on 2010-07-15 for ZOOM WHITENING LAMP AND PROCEDURE KIT ZM2550 manufactured by Discus Dental, Llc.

Event Text Entries

[1363516] The pt claims as a result of the procedure, that she experienced swelling of the upper and lower lip with a blister on the upper inside of the lip. The dentist ((b)(6)) prescribed peridex and warm salt water to use. The pt stated they went to the emergency room because, their lip was still swollen because (b)(6) "was no help. "
Patient Sequence No: 1, Text Type: D, B5


[8588720] The discus technical specialist for zoom chairside whitening ((b)(4)) investigated the incident and evaluated the process/procedure(s) that (b)(6) office used in whitening the pt's teeth. It was strongly recommended that the dfu for the whitening procedure be followed; up to the including using all materials provided; most specifically the isolation retractors. The dfu provided with the kit was reviewed to make sure it contained best practice techniques to avoid uv exposure to lips and gums as well as initial and follow-up training.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2032714-2010-00002
MDR Report Key1785048
Report Source04,05
Date Received2010-07-15
Date of Report2010-07-09
Date of Event2010-06-08
Date Facility Aware2010-06-09
Report Date2010-07-09
Date Reported to FDA2010-07-09
Date Mfgr Received2010-06-09
Device Manufacturer Date2007-01-18
Date Added to Maude2012-03-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street8550 HIGUERA ST
Manufacturer CityCULVER CITY CA 90232
Manufacturer CountryUS
Manufacturer Postal90232
Manufacturer Phone3108458336
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameZOOM WHITENING LAMP AND PROCEDURE KIT
Product CodeEEG
Date Received2010-07-15
Model NumberZM2550
Catalog NumberZM2550
Lot NumberNA
ID NumberNA
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDISCUS DENTAL, LLC
Manufacturer AddressCULVER CITY CA US


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2010-07-15

© 2026 FDA.report