BOND READY TO USE TTF-1 PRIMARY ANTIBODY PA0364

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2010-07-27 for BOND READY TO USE TTF-1 PRIMARY ANTIBODY PA0364 manufactured by Leica Biosystems Newcastle Ltd..

Event Text Entries

[1570344] The customer reported to leica microsystem false positive staining on paraffin sections of thymoma with ttf-1 (thyroid transcription factor-1) antibody. Lung and thyroid tumors are known to be positive with tff-1 antibodies but staining on thymomas has not previously been described. No misdiagnosis occurred in this case.
Patient Sequence No: 1, Text Type: D, B5


[8679077] Leica's investigation of this complaint has determined that the staining in thymoma is correct and not a false positive. However, the instructions for use only indicate that thyroid and lung tumors should be positive. Based on our current investigations, the instructions for use for this product will be updated to reflect the staining of other tissues. Customers will be informed of this change with a field correction.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004859032-2010-00001
MDR Report Key1786222
Report Source07
Date Received2010-07-27
Date of Report2010-06-29
Date of Event2010-06-29
Date Mfgr Received2010-06-29
Device Manufacturer Date2009-08-01
Date Added to Maude2011-02-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer StreetBALLIOL BUSINESS PARK BENTON LANE
Manufacturer CityNEWCASTLE UPON TYNE NE1 8EW
Manufacturer CountryUK
Manufacturer PostalNE1 8EW
Manufacturer Phone912150567
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBOND READY TO USE TTF-1 PRIMARY ANTIBODY
Generic NameIMMUNOHISTOCHEMISTRY REAGENT
Product CodeNJT
Date Received2010-07-27
Model NumberTTF-1
Catalog NumberPA0364
Lot Number08679
Device Expiration Date2012-07-29
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLEICA BIOSYSTEMS NEWCASTLE LTD.
Manufacturer AddressBALLIOL BUSINESS PARK BENTON LANE NEWCASTLE UPON TYNE NE128EW UK NE12 8EW


Patients

Patient NumberTreatmentOutcomeDate
10 2010-07-27

© 2026 FDA.report