ACCESS 2 IMMUNOASSAY SYSTEM N/A 81600N

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-08-05 for ACCESS 2 IMMUNOASSAY SYSTEM N/A 81600N manufactured by Beckman Coulter Inc..

Event Text Entries

[1626770] A customer contacted beckman coulter inc. (bci) in regards to erroneous elevated ckmb result generated by access 2 immunoassay analyzer. Patient sample was repeated on an alternate unit and normal ckmb result was obtained. The erroneous result was not reported out of the laboratory. Patient treatment was not impacted by this event.
Patient Sequence No: 1, Text Type: D, B5


[8677085] The sample was collected in li heparin plasma tubes with a gel separator and centrifuged for six minutes at 3000 rpm. Qc was within the lab's established ranges pre and post the event. The customer performed a routine system check on (b)(4) 2010 which failed the washed mean. Bci field service engineer (fse) cleaned and verified the unit is working properly. Fse did not note any issues with the unit. A clear root cause of this event can not be determined.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2122870-2010-00399
MDR Report Key1786665
Report Source05
Date Received2010-08-05
Date of Report2010-08-04
Date of Event2010-07-14
Date Mfgr Received2010-07-14
Device Manufacturer Date2001-05-03
Date Added to Maude2010-08-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BLVD
Manufacturer CityBREA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC
Manufacturer Street1000 LAKE HAZELTINE DRIVE
Manufacturer CityCHASKA MN 55318
Manufacturer CountryUS
Manufacturer Postal Code55318
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameACCESS 2 IMMUNOASSAY SYSTEM
Generic NameDISCRETE PHOTOMETRIC CHEMISTRY ANALYZER
Product CodeJHS
Date Received2010-08-05
Model NumberN/A
Catalog Number81600N
Lot NumberN/A
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address1000 LAKE HAZELTINE DRIVE CHASKA MN 55318 US 55318


Patients

Patient NumberTreatmentOutcomeDate
10 2010-08-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.