UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM N/A 973100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-08-06 for UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM N/A 973100 manufactured by Beckman Coulter Inc..

Event Text Entries

[17059638] A customer contacted beckman coulter inc. (bci) in regards ckmb result above the normal reference range for one patient sample generated by unicel dxi 800 access immunoassay analyzer. Patient sample was re-centrifuged and repeated on the same unit and alternate unit. The results obtained from both runs were within normal ckmb reference range. The erroneous results were not reported out of the laboratory. Patient treatment was not impacted by this event.
Patient Sequence No: 1, Text Type: D, B5


[17285891] The sample was collected in a pst tube and centrifuged for 10 minutes at 3400 rpm. The tube was not properly filled to the fill line. Two levels of qc are run every 8 hours and have been recovering with the customer's established ranges. A bci field service engineer (fse) performed preventive maintenance and verification testing which met the published specification. No hardware issues were detected that would have contributed to this event. A clear root cause of this event cannot be determined.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2122870-2010-00395
MDR Report Key1787607
Report Source05
Date Received2010-08-06
Date of Report2010-08-06
Date of Event2010-07-13
Date Mfgr Received2010-07-13
Device Manufacturer Date2007-08-30
Date Added to Maude2010-08-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BLVD
Manufacturer CityBREA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC
Manufacturer Street1000 LAKE HAZELTINE DRIVE
Manufacturer CityCHASKA MN 55318
Manufacturer CountryUS
Manufacturer Postal Code55318
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM
Generic NameDISCRETE PHOTOMETERIC CHEMISTRY ANALYZER
Product CodeJHS
Date Received2010-08-06
Model NumberN/A
Catalog Number973100
Lot NumberN/A
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address1000 LAKE HAZELTINE DRIVE CHASKA MN 55318 US 55318


Patients

Patient NumberTreatmentOutcomeDate
10 2010-08-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.