MTS INCUBATOR MTS9680

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2010-08-26 for MTS INCUBATOR MTS9680 manufactured by Microtyping Systems.

Event Text Entries

[19621595] The customer immediately took the incubator out of service. The incubator is under warranty and the customer is requesting repair by replacement. (b)(4)
Patient Sequence No: 1, Text Type: N, H10


[19685078] Customer reported that the mts incubator temperature read 44 degrees c during patient testing. The incubator temperature spec is 37 + / - 2 degrees c. The incubator was taken out of use. No erroneous results were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1056600-2010-00111
MDR Report Key1815542
Report Source05,06
Date Received2010-08-26
Date of Report2010-08-26
Date of Event2010-08-23
Date Mfgr Received2010-08-23
Device Manufacturer Date2006-05-24
Date Added to Maude2011-04-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. JENNIFER SPARROW
Manufacturer Street1001 ROUTE 202
Manufacturer CityRARITAN NJ 08869
Manufacturer CountryUS
Manufacturer Postal08869
Manufacturer Phone9082181000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMTS INCUBATOR
Generic NameINCUBATOR
Product CodeJTQ
Date Received2010-08-26
Catalog NumberMTS9680
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROTYPING SYSTEMS
Manufacturer Address1295 SOUTHWEST 29TH AVENUE POMPANO BEACH FL 33069 US 33069


Patients

Patient NumberTreatmentOutcomeDate
10 2010-08-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.