Event Text Entries

[17718468] Male pt had extensive avulsions and lacerations on his lower leg from getting caught up in a boat propeller. The wound was initially dressed with a towel, but was soaking through. There were 2 wounds approximately 8-10 inches long and approximately 4 inches wide, one on the medial side of the left lower leg, the other on the lateral side of the left lower leg. The mat tourniquet was applied approximately 3 inches above the knee. Upon tightening, the medic noticed that the venous bleeding had stopped, but the arterial bleeding was still present. The medic continued to tighten the tourniquet and as arterial bleeding was slowing down, the web strapping slid through the buckle. The medic tried to tighten the strap through the buckle again and attempted to tighten the tourniquet again but it happened again. At this point, the "(b)(6)" was landing so the medic packed the wound with trauma dressing. The (b)(6) nurse exposed and clamped off the arteries. The pt was ok. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[17982861] Still under investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9615387-2010-00011
MDR Report Key1825951
Report Source08
Date Received2010-08-31
Date of Report2010-08-31
Date of Event2010-08-01
Date Mfgr Received2010-08-17
Date Added to Maude2012-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street#7, 13511 CRESTWOOD PL
Manufacturer CityRICHMOND, BC V6V2E9
Manufacturer CountryCA
Manufacturer PostalV6V 2E9
Manufacturer Phone3037964
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMAT MECHANICAL ADVANTAGE TOURNIQUET
Product CodeGAX
Date Received2010-08-31
Model NumberMAT01
Lot NumberNI
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPYNG MEDICAL CORP.
Manufacturer AddressRICHMOND, BC V6V2E9 CA V6V 2E9


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2010-08-31