SPACELABS ULTRAVIEW SL COMMAND MODULE 91496

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-09-04 for SPACELABS ULTRAVIEW SL COMMAND MODULE 91496 manufactured by Spacelabs Healthcare.

Event Text Entries

[16837387] Spacelabs has received a report of significant noise observed on the ecg signal when other devices were energized in an operating room.
Patient Sequence No: 1, Text Type: D, B5


[16937991] Spacelabs was able to confirm the reported symptoms at the customer's site. Spacelabs has worked with the customer to optimize the user setting and adjust cabling to minimize the interference. No one has been injured as a result of this condition. The hospital is continuing to use the equipment. The spacelabs investigation is on-going. We will provide a supplemental report once our investigation is complete.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3023361-2010-00028
MDR Report Key1827539
Report Source05
Date Received2010-09-04
Date of Report2010-09-03
Date of Event2010-08-05
Date Mfgr Received2010-08-05
Date Added to Maude2011-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDAVID GERAGHTY, MANAGER
Manufacturer Street5150 220TH AVE SE
Manufacturer CityISSAQUAH WA 98029
Manufacturer CountryUS
Manufacturer Postal98029
Manufacturer Phone4256577200
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSPACELABS ULTRAVIEW SL COMMAND MODULE
Generic NameMULTIPARAMETER MODULE
Product CodeLOS
Date Received2010-09-04
Model Number91496
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSPACELABS HEALTHCARE
Manufacturer Address5150 220TH AVE SE ISSAQUAH WA 98027 US 98027


Patients

Patient NumberTreatmentOutcomeDate
10 2010-09-04

© 2026 FDA.report