SYNCHRON? CX3 DELTA CLINICAL SYSTEM 467501

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-09-13 for SYNCHRON? CX3 DELTA CLINICAL SYSTEM 467501 manufactured by Beckman Coulter Inc..

Event Text Entries

[17062035] A customer contacted beckman coulter inc. (bci) regarding a leak inside the flow cell compartment area and shift in carbon dioxide (co2) qc results generated by the synchron cx3 delta clinical system. No erroneous results were reported outside the laboratory and no operator exposure during this hardware failure. The customer indicated that there was no affects on patient treatment because the issue was caught by observing the qc shift results.
Patient Sequence No: 1, Text Type: D, B5


[17283437] A field service engineer (fse) visited and performed routine repairs. Fse discovered that the co2 acid port (line 57) was disconnected at the flow mixer. Reconnected the line, which resolved the leaking issue. Fse also replaced the co2 electrode and membrane and recalibrated co2. After repairs, qc met specifications and recovery was improved.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2010-00740
MDR Report Key1832484
Report Source05
Date Received2010-09-13
Date of Report2010-09-10
Date of Event2010-08-24
Date Mfgr Received2010-08-24
Device Manufacturer Date1997-05-22
Date Added to Maude2010-11-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS NORA ZEROUNIAN
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameSYNCHRON? CX3 DELTA CLINICAL SYSTEM
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeJJF
Date Received2010-09-13
Model NumberCX3 DELTA
Catalog Number467501
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S KRAEMER BOULEVARD BREA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2010-09-13

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.