UT69 TABLE MOUNTED RADIATION SHIELDS UT6902-US

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2010-09-09 for UT69 TABLE MOUNTED RADIATION SHIELDS UT6902-US manufactured by Mavig Gmbh.

Event Text Entries

[15555868] The ut69table shield has a u profile that covers the shields support bar and the profile edges were sharp causing paper cut type injuries on the installer. (b)(6) did a quality check and found out that the manufacture did not burr (sand and polish) the edges which caused this issue. (b)(6) had the manufacture fix the issue and we went through and replaced all the defective profiles with ones that were burred and sanded so they had a smooth edge and did not cause any more cuts. The manufacturer of the u profiles detected that the problem came from a non sufficiently trained employee, has since increased his quality and training process and adapted a laser processed production.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5017368
MDR Report Key1833208
Date Received2010-09-09
Date of Report2010-09-09
Date of Event2010-07-14
Date Added to Maude2010-09-20
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameUT69 TABLE MOUNTED RADIATION SHIELDS
Generic NameUT69
Product CodeKPY
Date Received2010-09-09
Model NumberUT6902-US
Catalog NumberUT6902-US
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerMAVIG GMBH
Manufacturer AddressSTAHLGRUBERRING 5 MUNICH, BAVARIA 81829 GM 81829


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2010-09-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.